《Comparative Study on the Effect of Extractant on the Detection of Imidaclothiz Residue in Tea》
Selection of detection wavelength The spectrum of imidaclothiz standard solution was analyzed with DAD diode array detector.It had the largest absorption at 270 nm,with less interference from impurity peaks.The peak areas at multiple wavelengths around270 nm were compared(Table 1),and the results showed that the peak area of imidaclothiz at 270 nm was the highest,and the spectrum of imidaclothiz standard is shown in Fig.2.