《Table Imaging Follow-up Results》
本系列图表出处文件名:随高清版一同展现
《Safety evaluation of the biodegradable polymer-based sirolimus-eluting stent with faster drug eluting rate and polymer absorption kinetics in complex patient and lesion subgroup》
Results Between December 2013 and August 2014,2,348patients were randomly assigned to treatment with Bu MA(n=1,174)or Excel SES(n=1,174).Among those patients,1,713 patients(858in Bu MA SES group and 855 in Excel SES group)were recognized in complex patient and lesion subgroup.At 2 years,the incidence of definite/probable stent thrombosis was significantly lower with Bu MA SES(0.7%vs 1.9%,P=0.03),which is mainly driven by lower subacute(>24 h to 30 days)stent thrombosis(0%vs 0.6%,P=0.03)compared with Excel SES.However,in the non-complex population,thrombosis event was similar with Bu MA or Excel SES(0.7%vs 0%,P=0.50).Myocardial infarction event rate in the complex patient and lesion subgroup was similar with Bu MA or Excel SES(5.7%vs 6.0%,P=0.79),as well as TLF(8.8%vs 7.5%,P=0.34).Major outcomes were shown in Table.
图表编号 | XD0027950200 严禁用于非法目的 |
---|---|
绘制时间 | 2018.08.01 |
作者 | Sida Jia、Changdong Guan、Jinqing Yuan、Runlin Gao、Bo Xu |
绘制单位 | Department of Cardiology, National Center for Cardiovascular Diseases and Fuwai Hospital,CAMS and PUMC、Department of Cardiology, National Center for Cardiovascular Diseases and Fuwai Hospital,CAMS and PUMC、Department of Cardiology, National Center for Car |
更多格式 | 高清、无水印(增值服务) |