In order to satisfy 14C measurement with the new developed 200kV single stage electrostatic accelerator mass spectrometer,the sample preparation processes for 14C were investigated.The graphitization experimental conditions for H2-Fe,Zn-Fe and Zn-H-Fe methods were optimized on self-designed 14C sample preparation system.The beam current of 12C-for all prepared samples can reach 15-40μA and its stability is better than0.2%.The 14C measurement shows that the lower ratio of(12CH-+13C-)/12 C-benefits the high sensitive and miniaturization device.The result also shows that the Zn-Fe method can depress the ratio of(12CH-+13C-)/12C-for the samples with the carbon content less than 1mg,while,for the carbon content among 1 mg to5mg,the Zn-Fe and Zn-H-Fe methods have the equivalent effect.Through above preparation methods,the yield of carbon can reach more than95%for the carbon content of 100μg-5mg.