《Effect of guide vane outlet angle on pump performance and impeller radial force in an axial-flow heart pump》


  1. 获取 高清版本忘记账户?点击这里登录
  1. 下载图表忘记账户?点击这里登录

Parameters of the heart pump at the design operational condition are as follow:Flow rate Qd=5 l/min,nominal head H=120mm Hg,rotational speed n=5000r/min and the rotational frequency of rotor f=n/60=83.33Hz.As shown in Fig.4,the pump assembly mainly includes blade,guide vane,inlet pipe,outlet pipe,magnetic loop,motor stator,position sensor,pump shell and wire,etc[12].The main geometric parameters of the test pump are listed in Table 1.